Radiation Detector
NaI(Tl) + boronpolyester sandwich crystal detector with in-house high voltage power supply, lead collimator
Gamma energy range
25 keV ... 2.5 MeV
Sampling time
0.5 s
Angle of detection
conical, ±30°
Temperature range
-25 °C ... +50 °C
|
Alarm thresholds
automatic with
- background compensation
- speed compensation
- vehicle shielding-effect compensation
Alarm level for gamma radiation
1.02 ... 2.3 times the current background level
MDA with 99% probability, 2 m distance, 4.5 s exposition time:
for Am-241: 3.1 MBq,
for Cs-137: 1.0 MBq,
for Co-60: 0.5 MBq.
|
MDA for neutron source:
2.0·106 [n/s]
(using Cf-252 or Am/Be source)
Construction
The detector is delivered with a stand with the required height for the inspection, or with wall-mountable design.
Airtight, environmental resistant housing.
|